Adjustments / Programs

Dose rate adjustment

WARNING

Prerequisite

The "Adjustment mode" parameter in the "Dose rate adjustment" parameter group must
be programmed to "Semi automatic".

Preparations

Switch the SIREMOBIL Compact on.
Connect the service PC to the serial interface of the SIREMOBIL Compact.
Start the service program.
Select the "Diagnostics..."menu.
Select the "Dose rate adjustment" menu in the Combo Box "Group 1".
- The subsequent display is visible on the service PC:
- AUTO REGULATION
- BRIGHTNESS DEVIATION
- CONTROL CURVE
- ADJUSTMENT
- BV-ZOOM STATUS
- DOSERATE
Insert 25mm AL and 2.1mm Cu filters at the radiation exit port of the Sirephos.
The grid remains on the I.I.
Correction factors:
Grid factor 23cm (9") I.I. and 17cm (7") I.I.:
Chamber correction factor for Dali and Nomex:
Dose rate to be set at the I.I. input screen:
23 (9") I.I. and 17cm (7") I.I.:
Siemens AG
Medical Engineering
X-ray radiation!
See Chapter 1, Safety Information.
Protect yourself against radiation exposure. Wear a lead apron.
indicates the current status of the automatic adjustment
(active / not active / done)
indicates the difference of the brightness actual value minus the
brightness nominal value
indicates the currently selected fluoroscopic curve
Indicates the operating control mode
(Generator / Camera Iris / Stop)
indicates the I.I. format (survey / zoom).
indicates the selected dose control level (LOW / MID / HIGH)
Only the MID dose control level is used
Register 5
RXR2-130.061.02
Rev. 03
1.5
1.06
220 nGy/s
Page 5 of 22
10.03
CS PS 24
5 - 5
5
5
5
SIREMOBIL Compact
System Manual