Fluoroscopy Footswitch Configuration - Siemens ARCADIS Avantic Start-Up Instructions

Hide thumbs Also See for ARCADIS Avantic:
Table of Contents

Advertisement

36
USA
Service mask
input in 1/10 min.
Europe
Service mask
input in 1/10 min.
Make country-specific changes and click <Save>.
Click <Home>.
Perform a system shutdown.
Switch on the system and wait for the system to boot.
NOTE
If a change is made to the fluoro time, check this function.

Fluoroscopy footswitch configuration.

NOTE
Program the footswitch as follows.
Start service program and log in as described under "Service Login".
Select Main System / Configuration / Main System / Footswitch CFC / right.
Save the changes with <Save>.
Click on <Home> button.
Perform a system shutdown.
Switch on the system and wait for the system to boot.
ARCADIS Avantic
"Warning time limit"
5 min.
50
5 min.
50
With the introduction of the Hibernation Mode, system configura-
tion changes are not applied if the system is switched off and then
switched on again using the power buttons on the monitor cart.
To correctly apply the modified data, perform a system shutdown.
Open the <Options> - <End Session> menu and click the "Shut-
down system" button. Immediately thereafter, press the Power off
button on the monitor cart.
The system shuts down and switches off automatically.
The footswitch is set at the factory so that fluoroscopy is per-
formed with the left pedal and the freely-selectable operating
modes such as digital radiography, SUB, etc., are performed with
the right pedal. The setting can be changed to meet customer
requirements.
SPR2-330.815.01.03.02
12.06
Specific System Settings
"Blocking time limit"
10 min. (5 min. + 5 min.)
Page 36 of 66