GE
RAFT
DIRECTION 5775469, R
Type CF Patient Leakage Current Test - ECG Leads
ECG lead 1
Tester
Tester Polarity
Neutral
Switch
Closed
Normal
Closed
Normal
Closed
Reversed
Closed
Reversed
Open
Normal
Open
Reversed
ECG lead 2
Tester
Tester Polarity
Neutral
Switch
Closed
Normal
Closed
Normal
Closed
Reversed
Closed
Reversed
Open
Normal
Open
Reversed
ECG lead 3
Tester
Tester Polarity
Neutral
Switch
Closed
Normal
Closed
Normal
Closed
Reversed
Closed
Reversed
Open
Normal
Open
Reversed
10-34
3
DRAFT (J
EVISION
Ground Switch
Closed
Open
Closed
Open
Closed
Closed
Ground Switch
Closed
Open
Closed
Open
Closed
Closed
Ground Switch
Closed
Open
Closed
Open
Closed
Closed
Section 10-9 - When There's Too Much Leakage Current...
19, 2018)
ULY
Max Value
Value
Allowed
Measured
0.01mA
0.05mA
0.01mA
0.05mA
0.05mA
0.05mA
Max Value
Value
Allowed
Measured
0.01mA
0.05mA
0.01mA
0.05mA
0.05mA
0.05mA
Max Value
Value
Allowed
Measured
0.01mA
0.05mA
0.01mA
0.05mA
0.05mA
0.05mA
V
™ P8/V
OLUSON
OLUSON
B
S
ASIC
ERVICE
OK?
Comments
OK?
Comments
OK?
Comments
™P6
M
ANUAL