GE Voluson P8 Basic Service Manual page 350

Hide thumbs Also See for Voluson P8:
Table of Contents

Advertisement

GE
RAFT
DIRECTION 5775469, R
9.) The "TetrisConsole - Console Tester" window is displayed.
NOTE:
For description of keys and their functions as well as the key code numbers refer to: Figure 5-6-2: OPIO
layout on page 5-35.
Task
DISCONNECT
GET VERSION
LED TEST
Hardkey Test
Encoder Test
Slider Test
Trackball Test
10.)Exit the Test program by pressing CLOSE.
7-58
3
DRAFT (J
EVISION
Figure 7-52 Console Tester
Disconnect <--> Reconnect
The Console version is shown in the Status window on the screen.
Select the LED TEST button.
All LEDs will be switched on/off.
First click: All LEDs changes to green.
Second click: All LEDs go off.
Third click: All LEDs changes to orange.
Forth click: : All LEDs go off.
The CLEAR LED'S button sets all LED's off. (The illumination of all hardkeys is switched off.)
Press each hardkey.
Indication (+ key code number) of each key is shown in the Status window on the screen.
Swivel the encoders (digipots) left and right and check the display in the Status window on the screen.
Slider position on the screen should change correctly.
Check movement of the mouse cursor.
Section 7-9 - User Interface Troubleshooting
19, 2018)
ULY
Actions and expected Results
V
™ P8/V
OLUSON
OLUSON
B
S
M
ASIC
ERVICE
Status Window
™P6
ANUAL

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Voluson p6Voluson p6 bt1857782185778216

Table of Contents