GE Voluson P8 Basic Service Manual page 529

Hide thumbs Also See for Voluson P8:
Table of Contents

Advertisement

GE
DRAFT
DIRECTION 5775469, R
Section 8-53
Replacement of FAN with Cable ASSY II
8-53-1
Manpower
One person, 30 minutes.
8-53-2
Tools
Philips screwdriver, Allen Wrench
8-53-3
Preparations
1.) Power Off/Shutdown the system as described in
2.) Pull out the NEST as described in
8-53-4
Removal Procedure
1.) Unscrew 4 screws and remove fan connector.
8-53-5
Installation Procedure
1.) Install the new parts in the reverse order of removal.
NOTE:
In case of installing FAN with Cable ASSY (GPN:5669586), replace screws to SCREW PH M4X30
(GPN:5176333) in FRU.
2.) Perform:
3
DRAFT (J
EVISION
FRU8-47: Replacement of FAN with Cable ASSY II - Functional Tests
Chapter 8 - Replacement Procedures
19, 2018)
ULY
4-3-2 on page
Section 8-28 "Replacement of the NEST" on page