Part 10 Technical Data; General - Maquet ROTAFLOW User Manual

Table of Contents

Advertisement

10
TECHNICAL DATA
10.1

GENERAL

10.1.1
SPEED
Range
Display resolution
Display accuracy
10.1.2
FLOW
Range
Display resolution
Display accuracy
Offset drift
Maximum flow deviation with an
error
10.1.3
AMBIENT CONDITIONS
Temperature
Relative humidity
(non-condensing)
Air pressure
Note
If storage and transportation exceed 15 weeks, comply with the ambient
conditions for operation.
ROTAFLOW with ICU Kit
|
0 ... 5000 rpm
1 rpm
±20 rpm
0 ... 9.99 l/min
0.01 l/min
From 0 to 1 l/min:
0.07 l/min + offset drift
From 1 to 9.99 l/min:
7% of current value + offset drift
Max. 0.05 l/min
+20%
Operation
Storage
+10 ... +40ºC
-18 ... +45ºC
15 ... 95%
10 ... 96%
66 ... 106 kPa
66 ... 106 kPa
ROTAFLOW with ICU Kit | User's Manual | English | 1.0
Copyright MAQUET Cardiopulmonary AG, Hirrlingen
10 Technical Data
|
Transport
-18 ... +55ºC
10 ... 96%
66 ... 106 kPa
| 53 |