5 Safety
ROTAFLOW with ICU Kit
| 42 |
|
6
Remove the ultrasonic contact cream from the ROTAFLOW Centrifugal Pump
with a lint-free cloth.
7
Apply ultrasonic contact cream to
both sides (left and right) around the
outlet of the ROTAFLOW Centrifugal
Pump. Both sides must be
completely covered with the
ultrasonic contact cream.
8
Insert the ROTAFLOW Centrifugal
Pump into the ROTAFLOW Drive.
Check that the ROTAFLOW
Centrifugal Pump is positioned under
the holder of the ROTAFLOW Drive.
9
Close the lock.
10
Calibrate the flow measurement to zero:
– Make sure that a speed of 0 rpm is selected.
– Press the "Zero adjustment" button for three seconds to calibrate flow
11
Set the speed to 1500 rpm to overcome the patient and line pressure.
12
Open the tube clamp.
ROTAFLOW with ICU Kit | User's Manual | English | 1.0
Copyright MAQUET Cardiopulmonary AG, Hirrlingen
|
measurement to zero.