DISCONNECTION FROM A HEART-LUNG MACHINE DURING
OPERATION
You can proceed as follows in an emergency:
Disconnection
Monitoring system
Monitoring system and power supply
Monitoring system, power supply and
physical
1
If level monitoring is selected for the ROTAFLOW System, override it on the
heart-lung machine.
2
Connect the level sensor to the "CLS" socket on the ROTAFLOW Console.
3
Switch off the pump switch on the console of the heart-lung machine.
The ROTAFLOW System automatically switches to battery operation.
4
Select "STAND AL" mode for the ROTAFLOW-System (
31 and
The ROTAFLOW System may stop.
5
Press the "Bubble override" button to reset the bubble sensor (briefly, not so as
to override the function).
6
If no level sensor is connected, press the "Level override" button for three
seconds. The button flashes.
The ROTAFLOW System starts again.
7
Connect the power cord to the power socket of the ROTAFLOW Console.
8
Switch on the mains power switch at the rear of the ROTAFLOW Console.
9
Remove the ROTAFLOW Console.
10
Release the securing mechanism of the protective plug cap on the underside of
the ROTAFLOW Console and close the protective plug cap.
[SIG!]
is still displayed afterwards, follow the safety
4.2 "Selecting settings", page 33).
ROTAFLOW with ICU Kit
|
5.5 "Using Emergency
Steps to be performed
1 – 6
1 – 8
1 – 11
ROTAFLOW with ICU Kit | User's Manual | English | 1.0