10 Technical Data
| 54 |
10.2

ROTAFLOW CONSOLE

Power supply
Stand-alone operation
Operation on heart-lung
machine
Batteries
Serial output interface
Measurements and weight
10.2.1
POSSIBLE SETTINGS
Status display
SET MODE
SET BRIG
SET FLIM
SET LLIM
SET HLIM
SET CLAM
ROTAFLOW with ICU Kit | User's Manual | English | 1.0
Copyright MAQUET Cardiopulmonary AG, Hirrlingen
ROTAFLOW with ICU Kit
|
Meaning
Mode
Brightness of display
Lower flow limit
Upper speed limit
Upper speed limit
Not used: Automatic or manual tube clamp
|
Mains power supply
Line voltage
(factory-set)
Fuse
Frequency
Power supply via heart-lung machine
Voltage
Current
Type
Voltage
Capacity
Charge/discharge cycles
Shelf life
Transfer rate
Protocol
Dimensions (length × width ×
height)
Weight
Possible values
ART, ART PULS, STAND AL, FREE
(
4.1 "Modes", page 31).
25%, 50%, 100%
Off,
0.1 ... 0.9 l/min (resolution 0.1 l/min),
1.0 ... 5.0 l/min (resolution 0.5 l/min)
Off,
500 ... 4500 rpm (resolution 500 rpm)
Off,
3000 ... 5000 rpm (resolution 500 rpm)
100/115/230/240 V AC
For 100/115 V: 2 A
For 230/240 V: 1 A
50/60 Hz
+24 V
Max. 8 A
NiCd
24 V
5 Ah
Min. 500
Max. 6 months
(under storage conditions)
9600 Baud
8 N 1
179 × 385 × 243 mm
Approx. 14.4 kg