7 Maintenance and service
| 46 |
7
MAINTENANCE AND SERVICE
7.1

MAINTENANCE

WARNING!
Check battery capacity at least every six months.
Batteries should be replaced by MAQUET-authorized personnel:
Disposal of batteries
Return batteries to MAQUET for environmentally friendly disposal.
Remove the ventilation slot collar of the ROTAFLOW Drive and clean the
ventilation holes regularly.
Maintenance and repairs may only be performed by personnel authorized by
MAQUET.
7.2

SERVICE

The device must be serviced at least every 1000 hours of operation or every
12 months by MAQUET or MAQUET-authorized service personnel.
Make sure that devices are returned suitably packaged to protect them from
damage during transportation. Enclose a description of the problem together with
the name, address and telephone number of a contact person.
ROTAFLOW with ICU Kit | User's Manual | English | 1.0
Copyright MAQUET Cardiopulmonary AG, Hirrlingen
ROTAFLOW with ICU Kit
|
Every two years
If they cannot be fully charged within 8.5 hours
If fully charged batteries fail to operate the system for at least 30 minutes.
|