  • Page 1 ALPHAMAXX 1133.02BX/FX Operating instructions Universal operating table GA113302GB11...
  • Page 2 1133.02BX Overview Head rest (e.g. 1130.64) Fastening screws for head rest Locking lever for upper back plate Back plate, removable 1133.31X0 Back plate for general surgery Connection point for removable back plates Lower back plate Bellows Potential equalization Leg plates Socket for mains cable Locking lever for leg plate abduction Mains cable 16...
  • Page 3: Table Of Contents

    1133.02BX Contents Important safety instructions ....................... 4 Important information ........................6 Important notes on these operating instructions ............... 6 Using the operating table in accordance with the intended purpose .......... 6 Intended purpose ........................6 Fundamental safety aspects ....................6 III.
  • Page 4: Important Safety Instructions

    Danger of damage! Take care to avoid collisions between accessories and the operating table during adjustment procedures. Consult MAQUET prior to using accessories produced by other manufacturers if they affect the longitudinal geometry of the table top. There is no explosion protection while recharging the batteries or the operating table is on mains power.
  • Page 5 I. Important safety instructions 1133.02BX Danger of injury! Before aligning the leg plates horizontally, check the patient’s position carefully, particularly when using knee crutches. Danger of crushing! Do not touch the side rails in the area of the joints when adjusting the operating table top! Danger of pinching! Do not reach under the table top or between the table top components during adjustment...
  • Page 6: Important Information

    2. Using the operating table in accordance with the intended purpose The operating tables distributed by MAQUET are to be used exclusively in the field of human medicine. The user must have been instructed in the proper operation of the table before using it. This familiarization session is confirmed in an acceptance log.
  • Page 7: General Description

    1133.02BX III. General description III. General description 1. General features In principle, the ALPHAMAXX 1133.02 operating table is suitable for the following surgical procedures: 〈 in the thorax and abdominal areas 〈 in the head area 〈 in gynaecology 〈...
  • Page 8: Control Of Functions

    III. General description 1133.02BX 2. Control of functions The electrohydraulic functions of the operating table can be operated using the following control elements: • Cable-connected hand control (see 2.1) • Override system = emergency operation (see 2.2) or optionally: • IR remote control (see 2.3) •...
  • Page 9: Cable-Connected Hand Control - Functions

    1133.02BX III. General description 2.1 Cable-connected hand control The cable-connected hand control is connected to the operating table at one of the two sockets 14 at the over- ride control panel (for simultaneous connection of ca- ble-connected hand control and foot switch). The elec- tric motor-powered adjustments of the operating table can be made immediately by pressing completely the corresponding function keys.
  • Page 10: Override System = Emergency Operation

    III. General description 1133.02BX z) 0-position Resets the lateral tilt, Trendelenburg/reverse Tren- delenburg, back plate and leg plate adjustments to the horizontal position. If the function key z is pressed, the operating table top will first reset the adjustments in the following order – lateral tilt, Tren- delenburg/reverse Trendelenburg, back plate.
  • Page 11 1133.02BX III. General description The functions of the override control panel are shown in the following: A) LED status indicators of operating table Note: The status indicators are only active when the ca- ble-connected hand control or IR remote control is switched on.
  • Page 12: Ir Remote Control (Optional)

    III. General description 1133.02BX DOWN Slope leg plate downward LOCK Operating table locked UNLOCK Operating table on castors, movable 2.3. IR remote control (optional) The operating table can be equipped with an IR remote control, as desired. The required charging station for the IR remote control is not included with the order no.
  • Page 13: Ir System Code

    1133.02BX III. General description 2.3.1 IR system code An IR remote control is clearly assigned to each oper- ating table through a code. The operating table has the same system code as the associated IR remote con- trol. You can see from the label at the operating table column whether the IR reception has been activated for this operating table.
  • Page 14: Ir Charging Station

    III. General description 1133.02BX 2.3.2 Charging station Recharging the IR remote control: • Connect the charging station to the mains. The green LED in the charging station indicates whether the required supply voltage is available or not. • Place the IR remote control in the charging station as follows: –...
  • Page 15: Special Functions

    Be sure to observe the Operating instructions for the foot switch. 2.6 Special functions If requested, the MAQUET service personnel can set the NEUROLOCK function at the operating table. This function avoids adjustments of the table position due unintentional actuation of the keys at the hand control.
  • Page 16: General Features

    III. General description 1133.02BX 3. Table top 3.1 General features The 1133.02 has been designed to meet the require- ments of various surgical disciplines. Selecting from a variety of modules lets you adapt the geometry of the table top to accommodate individual patient height. The table top is divided into a maximum of 6 sections, depending on the type of leg plate and upper back plate module selected:...
  • Page 17: Attaching The Back Plate Modules

    1133.02BX III. General description 3.3 Attaching the back plate modules 1) • First remove the pad 2) • Loosen the eccentric safety levers at the right and left (3); press the safety push buttons x (fig.1) to do so. If a back plate is already installed, hold it with one hand.
  • Page 18: Leg Plates

    III. General description 1133.02BX 3.5 Leg plates 1133.53 • To mount the split leg plates, press the push button 10 and insert the leg plate pins into the mounting holes up to the stop (Fig. 5). They are locked in position automatically as soon •...
  • Page 19: Accessories

    Accessories which are mounted on the side rails of the operating table provided for this purpose should also been drawn exclusively from the MAQUET line of ac- cessories. MAQUET attaches great importance to care- ful harmonization with the operating tables in our range.
  • Page 20: Motor Drive For The Base

    III. General description 1133.02BX 4. Motor drive for the base The table top versions 1133.02B2/F3 are equipped with a motor drive for the base. A table equipped with a mo- tor drive for the base starts moving slowly and slows down softly after releasing the function key (i or k).
  • Page 21: Use Of The Operating Table

    PE cable during surgery please order p/n 0200 403 4 at your local MAQUET representative. 2. Operation on battery power When operating the table with the integral batteries with...
  • Page 22: Operation On Mains Power

    IV. Use of the operating table 1133.02BX Note We recommend to recharge the batteries overnight, but at least every 2 days, even if “low charge level” of the rechargeable batteries has not been indicated yet. Regular recharging of the batteries will not only en- sure sufficient battery power available for the opera- tion programme, it will also serve to prolong the use- ful life of the rechargeable batteries.
  • Page 23: Patient Positioning

    V. Patient positioning 1133.02BX V. Patient positioning 1. General The 1133.02 modular universal operating table offers the possibility to use various modules as the upper back plate, thus providing the following advantages: The operating table top which can be used for various purposes, can be adapted to meet the special requirements of diverse surgical disciplines (modules see Chapter VIII).
  • Page 24 V. Patient positioning 1133.02BX Danger of injury! If the table top width is not sufficient for an ergonomic position, the patient may be injured. The table width extenders (1001.75A0/ 76A0) can be mounted on the side rails to ensure an ergonomic patient position.
  • Page 25: Positioning A Patient Weighing Up To 135 Kg (300 Lb)

    V. Patient positioning 1133.02BX 2. Positioning a patient weighing up to 135 kg (300 lb) 2.1 Approved accessories Danger of injury! The mounted accessories may not exceed the maximum permissible length specified in the sketch. For further accessories, please refer to the accessories list, see Chapter VIII. 2.2 Operating table unlocked (UNLOCK) Danger of tipping the operating table over! If the patient is in the wrong position, the operating...
  • Page 26 V. Patient positioning 1133.02BX 2.2 Operating table unlocked (UNLOCK) Fig. V-1 Maximum patient weight 135 kg (300 lb), UNLOCK GA113302GB11...
  • Page 27 V. Patient positioning 1133.02BX 2.3 Operating table locked LOCK Fig. V-2 Maximum patient weight 135 kg (300 lb), LOCK GA113302GB11...
  • Page 28: Positioning A Patient Weighing From 135 Kg Up To 225 Kg (300 Lb. Up To 500 Lb.)

    V. Patient positioning 1133.02BX 3. Positioning a patient weighing 135 kg up to 225 kg (300 lb - 500 lb) 3.1 Approved accessories Danger of injury! The mounted accessories may not exceed the maximum permissible length specified in the sketch. For further accessories, please refer to the accessories list, see Chapter VIII.
  • Page 29 V. Patient positioning 1133.02BX 3.2. Operating table unlocked (UNLOCK) Fig. V-3 Maximum patient weight 225 kg (500 lb), UNLOCK GA113302GB11...
  • Page 30 V. Patient positioning 1133.02BX 3.3 Operating table locked (LOCK) Fig. V-4 Maximum patient weight 225 kg (500 lb), LOCK GA113302GB11...
  • Page 31: Positioning A Patient Weighing From 225 Kg Up To 450 Kg (500 Lb. Up To 1000 Lb.)

    V. Patient positioning 1133.02BX 4. Positioning a patient weighing 225 kg up to 450 kg (500 lb - 1000 lb) 4.1 Approved accessories Danger of injury! The mounted accessories may not exceed the maximum permissible length specified in the sketch. For further accessories, please refer to the accessories list, see Chapter VIII.
  • Page 32: Care And Maintenance

    VI. Care and maintenance 1133.02BX VI. Care and maintenance Cover the operating table top with sterile sheets when it is not in use. 1. Cleaning Use a mild alkaline general detergent (soap suds) con- taining tensides and phosphates as its active cleaning agents for cleaning the operating table and its accesso- ries.
  • Page 33: Maintenance

    VI. Care and maintenance 1133.02BX 3. Maintenance The operating table will have to be serviced once a year by the MAQUET factory service department or by service technicians authorized by MAQUET to do so. Please adhere to prescribed maintenance intervals as otherwise guarantee coverage will lapse.
  • Page 34: Environmental Protection

    The materials used for packaging MAQUET products are environmentally friendly. They are made of untreated wood, cardboard, recyclable plastics or other reusable materials. If you wish, MAQUET can also dispose of them. MAQUET products: MAQUET will take back any used MAQUET products,...
  • Page 35: Technical Data

    VII. Technical data 1133.02BX VII. Technical data 1. Operating table adjustments Length without head rest: 1900 mm (74.8 inches) Length with head rest: 2196 mm (86.5 inches) Width: 530 mm (20.9 inches) Width across side rails: max. 580 mm (22.8 inches) Height (without pads): 594 - 1056 mm (23.4 –...
  • Page 36: List Of Accessories

    VIII. List of accessories 1133.02BX VIII. List of accessories Fig.: VIII-1 Accessories for patients weighing up to 135 kg (300 lb). For further accessories, please refer to the accessories list, see Chapter VIII. GA113302GB11...
  • Page 37 VIII. List of accessories 1133.02BX Fig.: VIII-2 Accessories for patients weighing up to 225 kg (500 lb). For further accessories, please refer to the accessories list, see Chapter VIII. GA113302GB11...
  • Page 38: Accessories For The Head Rest Interface

    VIII. List of accessories 1133.02BX Fig.: VIII-3 Accessories for patients weighing up to 450 kg (1000 lb) For further accessories, please refer to the accessories list, see Chapter VIII. 1. Accessories for the head rest interface 1.1 for patients weighing up to 135 kg (300lb) 1150.71X0 Head rest approved only in conjunction with the 1150.37X0 carbon-fibre back plate for neurosurgery...
  • Page 39 VIII. List of accessories 1133.02BX 1.2 for patients weighing up to 225 kg (500lb) 1130.52X0 Head rest, simple version 1130.53X0 Head rest with double articulation 1130.54B0 Connecting bracket 1002.07A0 Head plate adapter 1002.71A0 Horseshoe-shaped head rest 1002.73A0 Head rest, flat 1002.6200 Head rest, flat 1002.7000...
  • Page 40: Accessories For The Back Plate Interface

    VIII. List of accessories 1133.02BX 2. Accessories for back plate interface 2.1 for patients weighing up to 135 kg (300lb) 1150.37A0 Carbon-fibre back plate In addition, the accessories listed under section 2.2 are approved. 2.2 for patients weighing up to 225 kg (500lb) 1133.31B0 Back plate for general surgery, SFC padding, Europe 1133.32B0...
  • Page 41: Accessories For Attachment To The Side Rails

    VIII. List of accessories 1133.02BX 4. Accessories to be attached to the side rails 1001.62A0 Hand operating table, Europa 9092.3214 Padding 1001.28C0 Hand support 1001.44C0 Arm posturing device 300 mm 1001.44D0 Arm posturing device 450 mm 1001.45C0 Upper arm posturing plate 1001.4700 Leg restraint cuff 1001.56A0...
  • Page 42: Control Modules

    IR transmitter without charging station 1133.91A0 Amplified IR transmitter incl. charging station (230V) 1133.91L0 System codes numbers and letters 9801.6234 Cover for leg plate 1132.65A0 9082.1223 Control line with two plugs for foot switc 0200.4273 Use only spare parts approved by MAQUET. GA113302GB11...
  • Page 43 1133.02BX GA113302GB11...
  • Page 44 All notes and technical data contained in this documentation reflect the status at the publication date. Since we constantly strive to improve all MAQUET products, however, we reserve the right to adopt modifications at any time, without giving prior notice. There might therefore be minor differences between the illustrations in this manual and the product’s current design.