5 Safety
ROTAFLOW with ICU Kit
| 44 |
|
11
Bubble and level monitoring in "STAND AL" mode
Monitoring is performed by the internal bubble sensor of the ROTAFLOW Drive
and a level sensor, which is connected to the ROTAFLOW Console.
In "STAND AL" mode the pressure is not monitored.
5.5

USING EMERGENCY DRIVE

A ROTAFLOW System can be easily replaced with a spare ROTAFLOW System.
In emergencies, the ROTAFLOW Centrifugal Pump can be operated manually with
the ROTAFLOW Emergency Drive.
1
Close the arterial and the venous
tube clamp.
2
Remove the ROTAFLOW Centrifugal
Pump from the ROTAFLOW Drive.
Insert it in the ROTAFLOW
Emergency Drive.
3
Lift the handle and turn clockwise.
The speed display shows the speed
from 1500 rpm to 5000 rpm in steps
of 500 rpm.
Above 5000 rpm, all LEDs are lit.
4
When the speed is sufficiently high,
open both tube clamps.
Note
Turning counterclockwise is not possible.
ROTAFLOW with ICU Kit | User's Manual | English | 1.0
Copyright MAQUET Cardiopulmonary AG, Hirrlingen
Fit the protective cap to the connection socket of the heart lung machine.
|