Make sure that the on/off switch of the ROTAFLOW Console is switched off.
4
Connect the level sensor to the "CLS" socket on the ROTAFLOW Console.
5
Press the "Level override" button (briefly) to switch on level monitoring.
Test the level monitoring function (
page 35).
3.6

CONNECT TUBES

WARNING!
Only use tubes which are designed for perfusion and approved for this
system.
Connect 3/8" × 3/32" PVC or silicone tubes to the ROTAFLOW Centrifugal
Pump. Tubes with thinner walls might collapse.
Connect the pump tube in the direction of the arrows on the ROTAFLOW
Centrifugal Pump.
Only use ultrasonic contact cream, which is listed as an
"Accessories", page 47).
Check that the ROTAFLOW Centrifugal Pump is securely attached during
operation.
During longer procedures, it is possible that the ultrasonic contact cream
might dry out, which can impair the results of the flow meter. For this reason,
the ultrasonic contact cream must be renewed every 48 hours or as soon as
the error message
cream", page 41).
ROTAFLOW with ICU Kit
|
4.3.2 "Level monitoring: Performance test",
[SIG!]
appears (
5.3 "Replacing ultrasonic contact
3 Connection and Installation
|
accessory.