Part 9 Messages; Messages During Self-Test; Error Messages - Maquet ROTAFLOW User Manual

Table of Contents

Advertisement

9
MESSAGES
9.1

MESSAGES DURING SELF-TEST

Speed
display
–OK–
9.2

ERROR MESSAGES

9.2.1
IN SPEED DISPLAY
Message
⇐⇐
⇑⇑⇑⇑
⇓⇓⇓⇓
!!!!
9.2.2
IN SPEED AND FLOW DISPLAYS
Message
ERR OR
LOW BAT.
NO SEL
Flow
Status
display
display
EX.X
RFC XXXX
RFDMXXXX
RFDS– – – –
EX.X
PUMP OK
Acoustic
Pump
alarm
stop
-
-
Yes
No
Yes
No
Yes
Yes
Acoustic
Pump
alarm
stop
-
-
Yes
No
Yes
Yes
ROTAFLOW with ICU Kit
|
Meaning
[X.X]: Software version
[XXXX]: Operating hours since the ROTAFLOW
Console's last servicing
[XXXX]: Operating hours since the master
ROTAFLOW Drive's last servicing
Slave ROTAFLOW Drive not connected
Self-test completed without errors.
Cause/Error source
Recommendation to turn flow regulator to zero.
Speed > upper speed limit (HLIM)
Speed < lower speed limit (LLIM)
More than one
[RUN AWAY]
Switch off pump to clear error.
Cause/Error source
Error: The status display shows a further message.
The batteries are almost depleted.
voltage during battery operation", page 40
Console control module in the heart-lung machine
ROTAFLOW with ICU Kit | User's Manual | English | 1.0
Copyright MAQUET Cardiopulmonary AG, Hirrlingen
9 Messages
|
| 49 |
within 20 seconds
5.2.2 "Battery