1 General
ROTAFLOW with ICU Kit
| 10 |
|
WARNING!
ATTENTION!
1.2.1
CLAMPING THE TUBE
WARNING!
Do not open the tube clamp at a high speed to prevent magnetic uncoupling.
ROTAFLOW with ICU Kit | User's Manual | English | 1.0
Copyright MAQUET Cardiopulmonary AG, Hirrlingen
Do not use the system near escaping flammable or combustible gases.
During longer procedures, it is possible that the ultrasonic contact cream
might dry out, which can impair the results of the flow meter. For this reason,
the ultrasonic contact cream must be renewed every 48 hours or as soon as
the error message
[SIG!]
Do not use the ROTAFLOW System in suction mode.
The ROTAFLOW System should only be used with a valve or other adequate
systems or methods (e.g., two tube clamps) preventing a reverse flow.
If the pump stops during an operation, the blood flow will be interrupted and
supply to the patient will cease. Please ensure that the cause of the
interruption to the pump is remedied as quickly as possible and that the pump
is started up again promptly.
Make sure that there is no contact between high-frequency measuring
equipment and the measuring system or tubes, as this may cause the flow
indicator to malfunction.
Do not touch the patient and the external interfaces of the device at the same
time.
Do not touch the plugs of the ROTAFLOW Console as electrostatic charging
and moisture may cause damage.
Bear in mind that the ROTAFLOW Console weighs approx. 15 kg.
Switch off the mains power switch at the rear of the ROTAFLOW Console
before maintenance or cleaning.
Ask MAQUET about the most environmentally friendly method of disposing of
parts.
First of all disable the flow regulation (
Clamp the tube to prevent a reverse flow.
Immediately turn the flow regulator to zero to prevent hemolysis.
When opening the tube clamp, simultaneously increase the speed.
|
appears.
4.5 "Flow regulation", page 36).