WARNING! Risk of decannulation and mechanical damage
When changing the mode of transport, repositioning the patient or moving the
patient outside the mode of transport, there is a risk of decannulation and
mechanical damage caused by strain on the tubes or impacts.
The greatest care should thus be exercised when carrying out these measures.
Take care in confined spaces, such as doorways and elevators.
Do not allow tubes or cables to hang down.
Ensure that there is no strain on tubes or cables.
Avoid mechanical impacts and knocks.
Avoid kinking tubes or cables.
Do not drop any of the components.
WARNING! Risk of power failure
An inadequate power supply may put the ROTAFLOW Console out of service.
Please therefore note:
Always start work with the battery of the ROTAFLOW Console fully charged.
Use the power supply of the vehicle or hospital whenever possible so as to
conserve the battery.
Take the ROTAFLOW Emergency Drive with you when transporting the
patient. Keep it at the ready for stop-gap operation in the event of pump
failure.
ATTENTION!
All components must be installed so that they can be monitored and operated
by the user during transportation. There must also be sufficient space to use
the ROTAFLOW Emergency Drive.
Take at least two tube clamps with you when transporting the patient. Keep
them at the ready for clamping tubes in the event of leakage.
All components must be attached to the holders provided for this purpose and
properly secured (ROTAFLOW Console on Mobile RFC Holder, ROTAFLOW
Drive on Mobile ELS Holder).
For transportation in aircraft, the Mobile ELS Holder and Mobile RFC Holder
must be attached to the floor rails or to the ELS Base Plate and securely
fastened. If the ELS Base Plate is used, it must be attached to the aircraft floor
rails and securely fastened.
For transportation in ground vehicles, the Mobile ELS Holder and Mobile RFC
Holder must be attached to the ELS Wall Holder and securely fastened.
During transportation, ensure that the correct ambient conditions are
maintained for all components.
ROTAFLOW with ICU Kit
|
ROTAFLOW with ICU Kit | User's Manual | English | 1.0
Copyright MAQUET Cardiopulmonary AG, Hirrlingen
1 General
|
| 13 |