2 System Components
| 18 |
[9]
[10]
[11]
[12]
[13]
[14]
ROTAFLOW with ICU Kit | User's Manual | English | 1.0
Copyright MAQUET Cardiopulmonary AG, Hirrlingen
ROTAFLOW with ICU Kit
|
"Audio off" button
"Level override" button
(only in "STAND AL" mode)
"Bubble override" button
(only in "STAND AL" mode)
"LPM Mode" button
"Zero adjustment" button
"Clamp" button
|
The "Audio off" button is used to switch off the acoustic
alarm temporarily.
The button flashes when the acoustic alarm is
switched off.
The acoustic alarm is automatically switched back on
if a new alarm condition occurs.
The "Level override" button is used to switch level
monitoring off and on in "STAND AL" mode (
""STAND AL" mode: Level monitoring", page 37).
To switch level monitoring off, press the button for at
least 3 seconds.
The button flashes when level monitoring is switched
off.
The "Bubble override" button can be used to reset the
bubble sensor after a bubble intervention and switch
bubble monitoring on and off (
mode: Bubble monitoring", page 37).
To reset the bubble sensor after a bubble
intervention, press the button briefly (less than
3 seconds).
To switch bubble monitoring off, press the button for
at least 3 seconds.
The button flashes when bubble monitoring is
switched off.
Use the "LPM Mode" button to switch flow regulation
4.5 "Flow regulation", page 36 on and off.
The button lights up when flow regulation is switched
on.
The "Zero adjustment" button can be used to calibrate
flow measurement to zero (
Calibration", page 34).
You can use the "Clamp" button to confirm that suitable
systems or methods are available (e.g., two tube
clamps) to prevent a reverse flow (
ROTAFLOW Console, Self-test", page 29).