9 Messages
ROTAFLOW with ICU Kit
| 52 |
|
Message
STOP ???
STOP BUB
STOP LEV
STOP PRX
STOP-INP
TEMP B⇑⇑
TEMP C⇑⇑
TEMP M⇑⇑
UNEXPECT
Notes: Temperatures
ROTAFLOW with ICU Kit | User's Manual | English | 1.0
Copyright MAQUET Cardiopulmonary AG, Hirrlingen
Acoustic
alarm
Yes
Yes
Yes
Yes
Yes
No
No
No
Yes
If the temperature of the electronics exceeds 45°C, the ventilator switches to
full speed.
If the ROTAFLOW System stops because the temperature is too high, it
cannot be started again until the temperature has fallen.
No temperature messages should appear under normal operating conditions.
However, if this does happen, contact MAQUET Service.
|
Pump
Cause/Error source
stop
Yes
Stopped by an intervention not yet identified.
Possible causes:
Communication error if monitoring system has
been switched off
Immediately after reprogramming of ROTAFLOW
Console
Yes
Stopped by bubble monitoring
Yes
Stopped by level monitoring
Yes
Stopped by pressure monitoring
[X]: Pressure monitoring 1 to 4
Yes
Monitoring intervention
No
Temperature of battery > 50°C
No
Temperature of electronics > 60°C
No
Temperature of ROTAFLOW Drive > 50°C
Yes
Software