4 Operation
ROTAFLOW with ICU Kit
| 32 |
|
Display in "SET
MODE" menu
STAND AL
FREE
Reset bubble sensor before starting "STAND AL" mode
You must reset the bubble sensor after a bubble intervention. This also applies to
bubble interventions prior to the actual application (e.g., during priming).
To do so, press the "Bubble override" button briefly (less than 3 seconds).
ROTAFLOW with ICU Kit | User's Manual | English | 1.0
Copyright MAQUET Cardiopulmonary AG, Hirrlingen
|
Shown in status
display
SA
FRE
Meaning
Pump as stand-alone device.
Interventions using an internal bubble sensor
and a level sensor connected (
AL" mode: Interventions", page 36).
Pump without interventions or communication
with heart-lung machine.
This allows you to continue operation, for
example, in emergencies when the ROTAFLOW
System stops due to an error in the monitoring
system.
4.6 ""STAND