Rotaflow Drive; Rotaflow Emergency Drive - Maquet ROTAFLOW User Manual

Table of Contents

Advertisement

Status display
SET BTMP
SET S-AL
10.3

ROTAFLOW DRIVE

Coupling
Speed
Pressure
Flow meter
Bubble sensor
Pump stop
Motor
Voltage
Dimensions without holder
(diameter × length)
Holder radius of action
Weight
10.4

ROTAFLOW EMERGENCY DRIVE

Coupling
Gear
Speed display
Dimensions without holder
Holder radius of action
Weight
ROTAFLOW with ICU Kit
|
Meaning
Blood temperature
(to optimize flow
measurement)
Soft alarm
Magnetic
0 ... 5000 rpm
Max. 750 mmHg
Ultrasonic
Ultrasonic
Air bubbles > 5 mm (0.065 cm³)
Brushless, DC
+24 V
102 × 150 mm
Approx. 470 mm
Approx. 3.2 kg
Range
Resolution
Accuracy
Diameter × length
Length (handle folded in)
Possible values
15 ... 40°C (resolution 5°C)
On, Off (
User's Manual for heart-lung
machine)
Magnetic
1:39
1500 ... 5000 rpm
500 rpm
±250 rpm
102 × 270 mm
185 mm
Approx. 210 mm
Approx. 2.2 kg
ROTAFLOW with ICU Kit | User's Manual | English | 1.0
Copyright MAQUET Cardiopulmonary AG, Hirrlingen
10 Technical Data
|
| 55 |

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents