Hisense Covidien HME2C26 User Manual page 9

Table of Contents

Advertisement

Surge
Conducted interference
induced by RF field
Voltage dip
Voltage interruption
UT: Rated voltage. For example, 25/30 cycles mean 25 cycles at 50 Hz or 30 cycles at 60 Hz.
RF wireless devices are now widely applied in medical places where medical devices and/or systems
are used. When they are used in close proximity to medical devices and/or systems, they may affect the
basic safety and performance of medical devices and/or systems. The Hisense LCD has passed the
immunity test in the following table and meets the relevant requirements of IEC 00112: 201.
Customers and/or users must keep the following recommended minimum distances between RF
Frequency
Test
Band
Frequency
(MHz)
(MHz)
385
380-390
450
430-470
Important Safety Precautions
IEC 61000-4-5
IEC 61000-4-6
IEC 61000-4-11
IEC 61000-4-11
Recommended minimum separation distance
wireless communication devices and this Hisense LCD.
Service
TETRA400
GMRS 460 FRS
460
± 1 kV, 100 kHz, for signal input/output ports
Cable for cable: ± 0.5 kV, ± 1 kV
Ground cable: ± 0.5 kV, ± 1 kV, ± 2 kV
3 V under 0.15 MHz to 80 MHz
6 V for ISM and/or amateur RF bands between 0.15 MHz
80% AM at 1 kHz
0% UT: 0.5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°, 270°, and 315°
UT%: 1 cycle; 70% UT: 25/30 cycles
0% UT: 250/300 cycles
Modulation
Pulse
modulation 18
Hz
FM
± 5 kHz,
deviation
VII
and 80 MHz
0° sine phase
Maximum
Distance
power
(m)
(W)
1.8
0.3
2
0.3
Immunity
Test Level
27
28