Advertisement

Quick Links

RSAG2.025.6700SS A Version
Hisense website: www.hisense.com
User Guide
Read the User Guide thoroughly before
using Hisense LCD monitor and keep it
properly for future reference.
HMD2G21S

Advertisement

Table of Contents
loading

Summary of Contents for Hisense HMD2G21S

  • Page 1 HMD2G21S User Guide Read the User Guide thoroughly before using Hisense LCD monitor and keep it properly for future reference. RSAG2.025.6700SS A Version Hisense website: www.hisense.com...
  • Page 2: Preface

    Read the User Guide thoroughly before using the LCD monitor and keep it properly for future reference. Hisense shall not be liable for any accidents arising from violations of the safety precautions and use instructions of the User Guide.
  • Page 3: Safety Information And Precautions

     It is recommended that you use the specified accessories of the LCD monitor, which can be used only with the spare parts and accessories that are manufactured or specified by Hisense.  It takes about 30 minutes for the electronic components of the LCD monitor to achieve stable performance.
  • Page 4 To avoid hazards, do not place the LCD monitor on a soft pad. If the LCD monitor will not be used for a long time, disconnect its power cable. The LCD monitor contains hazardous high voltage inside, so do not disassemble it by yourself. If the LCD monitor is abnormal, contact Hisense or an authorized distributor.
  • Page 5: Table Of Contents

    Contents Preface ..............................2 Safety Information and Precautions ....................3 Contents .............................. 1 Overview ............................. 1 Features ............................2 Technical Specifications ......................3 Accessories ..........................4 Installation ............................5 Front and Rear Views of the LCD Monitor ................. 5 Front View of the LCD Monitor and Buttons ................6 Rear View of the LCD Monitor and Input Ports ................
  • Page 6: Features

    Overview Thank you for purchasing Hisense LCD monitor, which complies with the DICOM Part 14 standard and is applicable to DSA, MRI, DR, CR, CT, and PET medical imaging. It is intended for professionally trained medical staff and provides the image viewing and medical diagnostic functions.
  • Page 7: Technical Specifications

    Overview Technical Specifications Model HMD2G21S Name LCD monitor Size 21.3 inches (540 mm) Panel type Gray scale LCD panel (IPS) Resolution 1600 pixels (H) x 1200 pixels (V) Display area 432 mm (H) x 324 mm (V) Pixel pitch 0.270 mm (H) x 0.270 mm (V)
  • Page 8: Accessories

    The accessory list is for reference only. If any component is lost or damaged, promptly contact Hisense or the specified distributor. Keep the original package properly. If you need to return this product, be sure to use the original package. Do not place any heavy weight on top of the package during transportation.
  • Page 9: Installation

    Installation Front and Rear Views of the LCD Monitor Original packaging The monitor is lifted to the highest position Rotate 90° clockwise to change to vertical Rear view of the LCD monitor in vertical orientation orientation CAUTION  When adjusting the height of the monitor, hold the left and right sides of the monitor with both hands.
  • Page 10: Front View Of The Lcd Monitor And Buttons

    Installation Front View of the LCD Monitor and Buttons Note  For instructions on using the buttons, see the "Operational Instructions and Function Adjustment" section.
  • Page 11: Rear View Of The Lcd Monitor And Input Ports

    Installation Rear View of the LCD Monitor and Input Ports Port Function Power port Connected to the power adapter. Connected to a PC for LCD monitor debugging or parameter setting. USB1 (uplink port) Input: 5 V, 1 A. DP IN/DVI Connected to the digital output port on the graphics card of the PC.
  • Page 12: Hanging Arm Installation

    Installation Hanging Arm Installation Remove the base assembly to install the required hanging arm or base.  Follow the instructions in the User Guide when installing the hanging arm or base.  Before using the hanging bracket or base provided by another manufacturer, ensure that it meets the following requirements and complies with the VESA standard: Screw hole spacing: 100 mm x 100 mm It is solid enough to bear the weight of the monitor (excluding the base) and accessories, such...
  • Page 13: Cable Connection

    Installation Cable Connection The LCD monitor provides multiple connections to a PC and allows you to switch to any of the channels for display. The following figure shows the PC connection diagram. DVI port DP IN port DP OUT port Digital Digital Digital...
  • Page 14: Usb Connection

    Installation DP IN port DP IN port DP OUT port USB Connection Connect the LCD monitor to a PC by using the USB cable. The following figure shows the PC connection diagram. USB1 port USB port...
  • Page 15: Operational Instructions And Function Adjustment

    Operational Instructions and Function Adjustment Operational Instructions Operational Instructions  Buttons of the LCD monitor When the LCD monitor is in vertical orientation, seven adjustment buttons and a power indicator exist in the lower right corner of the monitor frame. The following table lists the functions of them. Button Function Light box button.
  • Page 16: Light Box

    Operational Instructions and Function Adjustment Mode Procedure: Custom Press the button on the LCD monitor to go to the Mode screen. Press to select a proper display mode. In Custom mode, press ▲/▼ to adjust backlight brightness. In DICOM200, DICOM250, DICOM300, DICOM350, DICOM400, DICOM450, DICOM500, DICOM600, or DICOM700 mode, press this button to switch to DICOM User mode.
  • Page 17: Menu Adjustment

    Operational Instructions and Function Adjustment Menu Adjustment Main Menu Display  After power-on, press to go to the User menu, and press to close the menu.  The following table lists the configurable items of the menu. First-level Main Menu Second-level Menu Remarks Menu...
  • Page 18: Setting And Adjustment

    Operational Instructions and Function Adjustment Display Settings User menu Setting and Adjustment Mode  Brightness Press the button on the LCD monitor to go to the User menu. Contrast  Press ▲/▼ to select the main menu item you want to adjust or set, Gamma and press to go to the next menu.
  • Page 19 Operational Instructions and Function Adjustment Mode  Select the display mode to be adjusted. The following table lists the initial values of all modes. Mode Brightness (cd/m Gamma Value Custom Gamma 2.2 DICOM200 DICOM DICOM250 DICOM DICOM300 DICOM DICOM350 DICOM DICOM400 DICOM DICOM450...
  • Page 20: User Menu - Power Saving Manager

    Operational Instructions and Function Adjustment Scale  Set Scale to adjust the image display effect. The options include Full Screen, Zoom In, and Normal. The default value is Full Screen. Option Definition The image is displayed in full screen mode. Full Screen The image may be deformed due to the difference between the image resolution and screen resolution.
  • Page 21 Operational Instructions and Function Adjustment Procedure: User menu Power Saving manager Press the button on the LCD monitor to go to the User Power menu. Ambient Sensor Press ▲/▼ to select the -g- icon, and press to go to the Power Saving manager menu on the right.
  • Page 22: User Menu - Calibration Settings

    Operational Instructions and Function Adjustment User Menu – Calibration Settings  The Calibration Settings menu provides the Auto Calibration and Auto Validation items. Calibration Settings Procedure: User menu Auto Calibration Press the button on the LCD monitor to go to the User menu. Auto Validation Press ▲/▼...
  • Page 23: User Menu - Monitor Settings

    Operational Instructions and Function Adjustment User Menu – Monitor Settings  In the power-on state, the Monitor Settings menu provides the Language, LOGO, Signal Input, and Monitor Info items. Procedure: Monitor Settings User menu Press the button on the LCD monitor to go to the Menu Location Upper Left User menu.
  • Page 24 Operational Instructions and Function Adjustment  The LCD monitor automatically enters sleep mode when it detects no output signals for several seconds, and the power indicator lights on in orange.  While the LCD monitor is in sleep mode, the system automatically wakes up display when detecting an input signal.
  • Page 25: Troubleshooting

    Troubleshooting Troubleshooting Before repair, check the following table to see if you can identify the cause of the fault. If the fault persists after you follow the instructions in the User Guide, you can consider repair. Fault Symptom Solution  Check whether the power plug is securely connected to the socket.
  • Page 26: Safety Statement

     Consult Hisense.  Use the power adapter and power cable provided by Hisense, and ensure that the LCD monitor is properly connected to ground.  Do not modify or retrofit the LCD monitor by yourself. Failure to comply with cause interference and damage to the LCD monitor.
  • Page 27: Environmental Protection Information

    Environmental Protection Information Environmental Protection Information  The LCD monitor meets the environmental protection requirements of the Administrative Measures for Limiting the Use of Hazardous Substances in Electrical and Electronic Products.  During its environmental-friendly service life, the LCD monitor does not cause leakage and precipitation of harmful substances and other problems affecting the health of users in the process of use, so it can be used safely.
  • Page 28: Maintenance Disclaimer

    User Guide but encounters equipment failure, parts damage, or monitor brightness not reaching the brightness value recommended in the User Guide, Hisense will repair or replace the product according to its own judgment.
  • Page 29: Appendix: List Of Terms

    Appendix: List of Terms  Digital Imaging and Communications in Medicine (DICOM) DICOM was jointly developed by American College of radiology (ACR) and National Electrical Manufacturers Association (NEMA). DICOM-compatible device connections can transmit medical images and data. DICOM Part 14 specifies the standard display functions of medical gray scale images.
  • Page 30 Qingdao Hisense Medical Equipment Co., Ltd. Address: North, 3F, Outsourcing Center of Software Park, No. 169, Songling Road, Laoshan District, Qingdao, P.R.China Hisense national customer service hotline: 4009940707 Hisense service supervision email: haixinfuwu@hisense.com...

Table of Contents