This device has been tested and proven to meet the requirements of GB4943.1-2011 Safety of
Information Technology Equipment, GB9254-2008 Information Technology Equipment - Radio
disturbance characteristics - Limits and method of measurement, GB17625.1-2012
Electromagnetic compatibility - Limits - Limits for harmonic current emissions (equipment input
current ≤ 16 A per phase), EMC: IEC/EN 60601-1-2: 2014, EN 55011/CISPR11
(MDSC-2226-MNA: Class A; MDSC-2226 LED, and MDSC-2226:DDI: Class B).
When the LCD monitor is used with a medical imaging system, ensure that the equipment
connected to the LCD monitor's signal I/O complies with medical equipment standards. All
external connections with other peripheral devices must conform to the requirements of Article 16
of IEC60601-1, Edition 3.1, or the requirements for medical electrical systems in Table BBB.201
in IEC 60601-1-1.
The electromagnetic waves emitted by power supply equipment may affect the use of the LCD
monitor and damage it. Install the LCD monitor in an environment free of electromagnetic wave
interference.
If the LCD monitor generates interference or reacts to interference, take one or more of the
following measures:
Adjust the position or direction of affected equipment.
Increase the distance between the LCD monitor and the affected equipment.
Consult Hisense.
Use the power adapter and power cable provided by Hisense, and ensure that the LCD monitor is
properly connected to ground.
Do not modify or retrofit the LCD monitor by yourself. Failure to comply with cause interference
and damage to the LCD monitor.
Equipment classification
Anti-lightning class: Class I
operating mode: continuous
Protection level: IP45 for the front side only; IP22 for other sides
Medical device category (MDD 93/42/EEC or MDR 2017/745): Class I
EMC class (IEC60601-1-2): Group I Class B

Safety Information

32