12. INSTRUCTION FOR 940ML CORE

The volume for TCF-32 CONTINUOUS FLOW ROTOR in which 940ML CORE is used is
940ml.
940ML CORE is good for separating the sample which contains larger volume of pellet.
12.1 SPECIFICATIONS
Max. speed
Max. centrifugal force
K factor
Total capacity
Material
Weight of core
12.2 USAGE
(1)
Mount the core, titanium sleeve and
blades.
Make sure the three types of O-ring
are free from scratches.
Apply a thin coat of vacuum grease
on them before installation.
Put the blades on the pins on the
titanium sleeve.
(2)
Operation :
Prepare for sipping the TCF-32 rotor after the same manner of TCF-32 INSTRUCTION
MANUAL.
Caution :
The volume of 940ML core is differ from the standard core.
Determine the flow in rate to be separated referring to the following usage.
32,000 RPM
101,685 x g
106
940 ml
core: NORYL
titanium sleeve: Titanium
blade: NORYL
O-ring: Nitryle rubber
1,380 g
-44-
O-ring
(1AP3)
Core
O-ring
(1AG110)
O-ring
(1AG100)
Titanium
sleeve
Blade