4
Mount the rotor on the Rotor base, put on the
cover, and tighten by the furnished handle.
¡ Tighten the cover above the match point (see
right figure) of the match marks of rotor and
cover.
5
Insert the shaft (D) assembly
into the hole located at the
center of the cover.
¡ Make sure the O-rings are
free from scratches, and
apply a thin coat of
vacuum grease on them.
¡ The shaft (D) can be
introduced 15 mm in the
cover. Put it in until it is
held against.
¡ Two O-rings are used, on
the top and bottom ends of
the shaft (D).
6
Sufficiently tighten the swageloknut until the
clutch of the torque wrench functions.
¡ It functions at 5 N?m { 51kgf?cm} .
Match mark
Shaft (D)
O-ring
(1AS5)
Nut
O-ring
stopper
Nut
O-ring
stopper
O-ring
O-ring
(1AP4)
(1AP4)