2
Mount the core, titanium sleeve and blades.
Make sure the three types of O-ring are free
from scratches. Apply a thin coat of vacuum
grease on them before installation.
Put the blades on the pins on titanium sleeve.
Position the sample outlet hole in the core and
the titanium sleeve in relation to each other as
shown below.
3
Introduce in the rotor the assembly of core,
titanium sleeve and blades.
Sample outlet hole
Blade
Sample inlet hole
-13-
O-ring
(1AF3)
Core
O-ring
(1AG145)
O-ring
(1AP135)
Titanium
sleeve