4
Mount the rotor on the Rotor base, put on the
cover, and tighten by the furnished handle.
Tighten the cover above the match point (see
right figure) of the match marks of rotor and
cover.
5
Insert the shaft (D) assembly
into the hole located at the
center of the cover.
Make sure the O-rings are
free from scratches, and
apply a thin coat of
vacuum grease on them.
The shaft (D) can be
introduced 15 mm in the
cover. Put it in until it is
held against.
Two O-rings are used, on
the top and bottom ends of
the shaft (D).
6
Sufficiently tighten the swageloknut until the
clutch of the torque wrench functions.
It functions at 5 N・m {51kgf・cm}.
Match mark
Shaft (D)
O-ring
(1AS5)
Nut
O-ring
stopper
Nut
O-ring
stopper
O-ring
O-ring
(1AP4)
(1AP4)
-14-