Turn on the switch on the low temperature circulating water cooling bath, and then make
sure that cooling water circulates. Turn on the WATER and DRAIN switches on the lubricating
unit. Waste oil will be sucked. If the TROUBLE lamp for WATER is lit and the buzzer sounds,
water is not flowing or the water pressure is too low.
Inspect the entire seal assembly to check that there is no leakage of cooling water
17
Turn the screw socket (B) of the seal body
assembly to adjust its height to the position of the
red line on the pipe.
The upper surface of manifold (A) should be
positioned so that some of the red line is still
visible.
If the red line is invisible even by tightening the
screw socket (B) fully, the screw socket (A) may
not have been tightened sufficiently. In such a
case, repeat from 4.2.2. 14.
18
The assembled status is illustrated on the right.
-21-
Manifold (B)
Red line
Manifold (A)
Screw socket (B)