Mount the core, titanium sleeve and blades.
¡ Make sure the three types of O-ring are f r ee
from scratches. Apply a thin coat of vacuum
grease on them before installation.
¡ Put the blades on the pins on titanium sleeve.
¡ Position the sample outlet hole in the core and
the titanium sleeve in relation to each other as
shown below.
3
Introduce in the rotor the assembly of core,
titanium sleeve and blades.
Sample outlet hole
Blade
Sample inlet hole
-13-
O-ring
(1AF3)
Core
O-ring
(1AG145)
O-ring
(1AP135)
Titanium
sleeve