Fig. 9 Assembly of seal body
Assembly, collar and
Screw sockets
3.3 Lubricating unit assembly
The lubricating unit assembly is used when operating this continuous flow rotor.
The specification and operation procedure of this unit are as follows.
(1) Specification
Item
1. Lubricating oil supply
flow rate
2. Drain suction
pressure
3. Malfunction detection
4. Malfunction warning
5. The container
6. Noise
7. Input power
8. Size
9. Weight
Seal body
ass'y
Screw
socket (B)
Collar
Screw
socket (A)
Specification
80 ± 20 ml/min
81 kPa (The pump simple
substance pressure).
l Lubricating oil flow detection
(The pressure detection
formula)
l Cooling water flow detection
(The pressure detection
formula)
There is a warning in the
electronic buzzer.
The glass bottle of 500 ml
For lubricating oil and the waste
oil
60 dB (A scale)
l Input power :
AC 100-240V, 50/60Hz
l Inner supply voltage :
DC 12V, 4.17A max.
273 X 318 X 219
10.8 kg
-8-
Fig. 10 Seal body assembly
Remarks
l For the seal attachment
assembly lubrication
l For the DRAIN collection
l It is an extraordinary caution
when Lubricating oil doesn't
flow.
l It is an extraordinary caution
when cooling water doesn't
flow.
l 1 m of the front
l It uses AC adapter