1. PREFACE

We have developed a continuous flow rotor which permits sizable processing with samples
allowed to flow continuously.
Utilize this continuous flow rotor of high performance and reliability for your studies or
processing

2. SPECIFICATIONS

Table 1 shows the specifications of this continuous flow rotor.
Type of Rotor
Max. speed
Max. centrifugal force
K factor
Total capacity
Maximum flow rate
Sample pH range
Maximum Dia. of rotor
Supply power for lubricating unit
NOTES
For operation of Model TCF-32, the following apparatus are required besides the standard
configurations of Model TCF-32 and Preparative Ultracentrifuge, which have to be prepared for
installation of Model TCF-32 at customer's site.
1. Sample Loading Pump
Specifications
Flow rate: Has to be as fast as operating flow rate required for separation of samples.
Exhaust pressure: More than 1.5kg/cm
Model without pulsating motion is recommended.
2. Low Temperature Circulating Water Cooling Bath
Cooling capacity: More than 300 Kcal/hr (water temp. at 5°C)
(Refrigerator: as strong as 300W)
Circulating Pump
Flow rate: More than 1 l/min. (pressure: 0.5-1 kg/cm
3. Density Gradient Pump
Required for density gradient separation.
Table 1 Specifications
TCF-32
32,000 RPM
101,685 X g
430 ml
150 ml/min (9 l/hr)
PH 4~10
24 cm
AC 100-240V, 1A (50/60Hz)
2
.
2
)
-3-