Preparing a UTREL30-WiFi for first use
Once an alarm has triggered, the alarm indicator ( ) remains shown until the alarm is
cleared (see
Clearing an Alarm on
marker
remains shown until midnight, then it turns off and the marker for the
previous day is shown (
previous day. When midnight passes next, this marker will move to
Alarm Re-triggering
A note on alarm re-triggering:
delay time (but not any others) resets to zero and alarm processing starts again.
The alarm processing for all other alarm delays is not affected. Therefore,
accumulative or consecutive alarms will re-trigger, if the alarm conditions are met
again, and the Alarm Indicator ( ) and the Day Alarm Marker (
even if any previous alarm was cleared. Clearing an Alarm does not reset any of
the delay values.
UTREL30-WiFi User Guide
page 52) or the unit is reconfigured. The day alarm
) to indicate the alarm was registered against what is now the
As soon as an alarm is triggered, the corresponding
22
and so on.
) will be shown,