Kipp & Zonen BREWER MK IV Instruction Manual page 87

Table of Contents

Advertisement

UVRJJJYY.nnn
First column: wavelength in Angstrom.
Second column is responsivity in counts/mW/m
ZSFJJJYY.NNN (ZSFVAL)- ZENITH SKY CONSTANTS
Zenith Sky constants are used in the ZS ozone calculations, and are Location/Brewer dependent.
They are derived by making a comparison of near simultaneous DS and ZS measurements over a
wide range of mu and ozone values (usually many months).
The values supplied in ZSFVAL.nnn are for a Brewer #035 operating in Toronto, Canada, and
should produce results which are accurate to within 5% or so. These values can be used until a new
set can be derived for the new site.
Example
-0.006400
-0.019680
0.016540
0.123076
0.281095
-0.060974
-0.486491
0.458119
-0.044107
84
Name
Coefficient #1
Coefficient #2
Coefficient #3
Coefficient #4
Coefficient #5
Coefficient #6
Coefficient #7
Coefficient #8
Coefficient #9