Kipp & Zonen BREWER MK IV Instruction Manual page 79

Table of Contents

Advertisement

A/D VALUES:
The AP (Analog Printout) command generates the printout shown below. A/D values are not stored
in files (except HV and +5v and SL current), but they are very important in assessing the health of
the Brewer, so they are included here for reference.
When the AP printout is displayed on the screen only the first 18 entries are shown (channels 0 to
17). In order to see the remaining 6 channels the file must be printed to a line printer or to disk.
Channels 18, 19, 22 and 23 are not currently used and the values shown are not useful.
A/D Values for nov 02/99 at 17:05:59 for instrument number 173
Channel#
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
76
Name
PMT temp (deg C)
Fan temp (deg C)
Base temp (deg C)
H.T. voltage (V)
+12V power supply (V)
+ 5V power supply (V)
-12V power supply (V)
+24V power supply (V)
Rate meter (V)
Below Spectro temp (C)
Window area temp (C)
External temp (deg C)
+ 5V ss (V)
- 5V ss (V)
Std lamp current (A)
Std lamp voltage (V)
Mer lamp current (A)
Mer lamp voltage (V)
External 1 (V)
External 2 (V)
Relative humidity (%)
Moisture (g/m3)
External 4 (V)
External 5 (V)
MKIV OPERATOR'S MANUAL
Value
(Lamps off)
(Lamps on)
22.99
22.04
23.56
1490.97
1490.97
12.49
5.00
-11.79
24.48
0.00
22.23
22.52
23.84
5.01
-4.95
0.00
0.00
0.01
0.00
0.06
0.04
0.00
0.00
0.04
0.05
Value
22.99
21.95
23.56
12.49
4.99
-11.79
24.48
0.00
22.23
22.52
23.84
5.01
-4.95
1.53
9.58
0.37
9.40
1.08
0.34
0.00
0.00
0.05
0.05