Kipp & Zonen BREWER MK IV Instruction Manual page 84

Table of Contents

Advertisement

16-10-200716-10-2007
45
.2
46
64
47
256
48
0
49
64
50
0
51
2232
52
June 1/99
LAMPLLL.IRX - LAMP IRRADIANCE FILE
Lamp Irradiance files, generated at the factory, tabulate the UV irradiance of
First row - lamp identification number
Second row - distance between lamp filament and Teflon diffuser
Now the file splits into 2 columns:
First column - wavelength in Angstroms from 2865A to 3630A in 0.6nm Increments.
Second column - irradiance in milli watts per square meter.
OP_ST.NNN - OPERATING STATE FILE
The operating state file controls many of the operating parameters of the Brewer. The left column
lists the actual value written in the OP_ST.nnn file. The middle column is the BASIC variable name
used in the Brewer software to contain this value, and the right column is a description of the value's
meaning.
Sample
1
162
2
C:\bdata\
3
icf25899
4
zsf32098
5
dcf14799
6
uvr09899
7
18
8
09
9
99
10 Saskatoon LO$
11 52.108
12 106.713
13 960
14 3.54
15 810
16 -12
17 14666
18 1
19 1
20 1
21 1
22 1
23 0
Computer buffer delay (larger numbers for faster computers)
NO
filter wheel #1 position
2
Ozone filterwheel #1 position
Filterwheel #2 position
UV filterwheel #2 position
Steps from zenith sensor to the hard stop
Zenith UV position
Release Date
SW
Explanation
Variable
NO$
Brewer ID number
DD$
Data Directory
ICF$
instrument constants file
ZSF$
zenith sky coefficients file
DCF$
dispersion constants file
UVR$
UV response file
DA$
month Brewer was last in operation
MO$
Month Brewer was last in operation
YE$
Year Brewer was last in operation
Location of Brewer instrument
L1$
Latitude of instrument
L2$
Longitude of instrument
L3$
Average climatic station pressure (millibars)
TE$
Voltage representation of Brewer temperature
NC%
Azimuth north correction
HC%
Zenith horizon correction
SR%
Azimuth steps per revolution
Q1%
Zenith drive motor
Q2%
Azimuth drive motor
Q3%
Iris drive motor
Q4%
Filterwheel #1 drive motor
Q5%
Filterwheel #2 drive motor
Q6%
Clock board
MKIV OPERATOR'S MANUAL
APPENDIX B CONFIGURATION FILES
the UV test lamps.
81