Kipp & Zonen BREWER MK IV Instruction Manual page 119

Table of Contents

Advertisement

Example
R; O
0
If the above had been called before any other R command then a single zero is returned
indicating that there were no scans taken.
R,2,4,4;O[sent to Brewer]
5638,
4996,
Sample and accumulate the light intensities for slitmask positions 2, 3,4,4,3 and 2.
Repeat this sequence 4 times. In this example the O command is used to display
example results. The returned counts correspond to slitmask positions 2, 3 and 4
respectively.
R; O[sent to Brewer]
5549,
4989,
In the above, the R parameters had already been specified as 2, 4, 4.
The new values returned correspond to a new run of the R command as specified.
12. S
Report the status of the most recent run command. This command requires
configuration variables.
Syntax
S
Response
<p1>,<p2>,<count>,<p3>,0,<interrupted>,
Each number returned in the above string is in the form of a decimal value using ASCII
characters. Each number occupies exactly four characters and the value is right-justified
and padded with spaces. The last value in the list always has a comma. The number of
parameters returned is limited only by the string length.
Parameters
The <pi> are the corresponding parameter from the most recent R command. The
remaining parameters are given in the following table.
Response field
<count>
<interrupted>
116
Slitmask position
0
1
2
3
4
5
6
7
54886[returned by Brewer]
54880[returned by Brewer]
format
Meaning
0..<p3>
Number of cycles completed
Boolean
on if the Run command was interrupted (by a break)
MKIV OPERATOR'S MANUAL
meaning
Hg calibration: 302.1 nm
Dark Count
λ1 306.2 nm
λ2 310.0 nm
λ3 313.5 nm
λ4 316.8 nm
λ5 320.0 nm
λ2 & λ4 for Deadtime test