UMKSETUP.NNN - UMKEHR SETUP FILE
#
Example
1
SXUNI4M.DAT
2
STDTABS.DAT
3
O3TABLE.DAT
4
CQMS.DAT
5
4.1099 2.3155 1.5600 0.8637
6
0.6760 0.3187 0.1490 0.0776 .. ozone coefficients for #039 STN TMO
7
3
6
11 12
8
999
1
9
0
... debug file
1 - 4. Tables used in the calculation of Umkehr profiles.
5 - 6. Ozone coefficients
7. This line has 7 parameters:
1 Minimum starting zenith angle
2 Maximum starting zenith angle
3 Minimum ending zenith angle
4 Maximum ending zenith angle
5 Output control flag
6 Ozone scaling factor (used with Dobson data)
7 Surface pressure of the station (in mmHg)
8. This line has 4 parameters:
1 Station identifying number (3 digits)
2 Parameter no longer used (set to 1)
3 Latitude of station - used for first guess
4 Output control flag (set to 0)
9. Flag for debugging purposes only.
UVRJJJYY.NNN
First column: wavelength in Angstrom.
Second column is responsivity in counts/mW/m
82
Q7%
A/D board
Q8%
UVB port
Q9%
Filterwheel #3 drive motor
Q10%
New or old temperature circuit. Set to 1 for a new
temperature circuit and 0 for an old circuit.
Q11%
Second film polarizer
Q12%
Set to 1 to enable NOBREW operation
Q13%
Wide HG slit present. Always set this to 1.
Q14%
New Brewer electronics board
Q15%
Humidity Sensor
DI$
Schedule or menu indicator
MDD$
Mode
SK$
Schedule name
1
1.0 812.0
34.838 0
MKIV OPERATOR'S MANUAL
2
/nm.