16-10-200716-10-2007
Parameters
The following table gives the permissible values for <m>.
<m>
Motor
1
Zenith prism
2
Azimuth Tracker
3
Iris
4
Filterwheel 1
5
Filterwheel 2
6
Filterwheel 3
9
Micrometer 2
10
Micrometer 1
11
Slitmask 1
12
Slitmask 2
13
Zenith Tracker
Example
I,4
The motor used by Filterwheel 1 is initialized.
5. LOGENTRY
Report the next entry in the log
Syntax
LOGENTRY
Response
A character string identifying a problem. This string has the form
YYYY DDD HH:MM:SS <text>
where the text may be (for example)
1995 033 14:22:03 All log entries reported
Entry
LOGENTRY
1996 302 13:10:02 Warm reset requested. Tepid reset generated (RAM was
corrupt).
Reading the oldest unreported log entry;
LOGSTART
LOGENTRY
1996 302 13:10:07 Motor #8: IIC communications error
LOGENTRY
1996 302 13:10:14 MUGWUMP: symbol not found
reading the oldest log entry;
LOGFINISH
LOGENTRY
1996 302 13:10:22 All log items reported