6.
A plot of the dark count and wavelength 1 intensity values vs the high voltage should be
compared to the plot in the Final Test Record.
The wavelength 1 data should exhibit a region of minimum slope (plateau).
The optimum setting for the high voltage is at the beginning of the plateau. The ratio is
determined by deviding the counts of wavelength 1 by the square root of the dark count.
If the new established optimum high voltage value deviates approximately 50 Volts or more
from the orignial setting then it is desired to change the current setting. See the HVSET
explanation on how to adjust the high voltage.
Changing the high voltage for the photomuliplier may effect the ozon
WARNING:
calibration, but will change the UV calibration for sure. The best time for
adjusting the high voltage is therefore during re-calibration of the Brewer
instrument.
HVSET: Set PMT High Voltage
HVSET is used to set the PMT High Voltage as determined from the HV test.
This test requires access to the High voltage module and the front of the Main Electronics Circuit
Board.
Prior to starting the test, locate shorting plug J2 and potentiometer R4 on the High Voltage Module,
and test point E16 on the Main Electronics Board.
Method:
1. If not already done so, then turn off Brewer power and on the high voltage module move jumper
J2 from Pins 2-3 to Pins 1-2, and restore Brewer power.
2. Enter the HVSET command and enter the desired HV value at the prompt.
3. Measure the voltage at E16 on the Main Electronics PCB and write the value down.
4. Turn off Brewer power and set jumper J2 back to pins 2-3, and restore power.
5. Monitor E16 and adjust R4 on the High Voltage Module until E16 is set to the voltage as recorded
in step 3.
6. Confirm the setting with an AP command.
4
16883
5
27775
7
40973
6
55779
10
71077
9
86191
8
99579
17
111900
9
122532
11
131484
14
138978
18
144932
17
149962
15
153943
18
157825
19
160926
19
163407
22
165796
19
168082
24
170292
25
172626
26
175036
29
177199
MKIV OPERATOR'S MANUAL
APPENDIX F FACTORY TESTS
8442
869.5
12421
879.5
15486
891.5
22772
901.6
22477
911.6
28730
923.6
35206
933.6
27140
943.7
40844
955.7
39644
965.7
37143
975.7
34161
985.7
36371
995.8
39748
1007.8
37200
1017.8
36919
1027.8
37488
1037.9
35348
1049.9
38561
1059.9
34761
1069.9
34525
1082.0
34327
1092.0
32905
1102.0
99

Advertisement

Table of Contents
loading

Table of Contents