Technical data
DC/DC converter

Gas supply

1) NIST = Non-Interchangeable Screw-Threaded
2) DISS = Diameter Index Safety Systems
3) N-F = Norme française
148
IP classification
Temperature range
Input
Output
O
supply pressure
2
Supply gas
Connector for O
supply
2
Gas cylinders and pres-
sure reducers
Pressure reducer
Gas consumption for
internal control
Accuracy of gas
consumption indication
IP42
–20 °C to +50 °C
12/24/28 VDC; 5 A/2.5 A/2.1 A
19 V ±0.5 V/2.6 A
The intended use of the DC/DC converter is in vehi-
cles.
From a central gas supply system or from an
O
cylinder.
2
270 Pa to 600 kPa at 100 L/min
Medical oxygen
either:
1)
NIST
acc. to EN 739/ISO5359 or
2)
DISS
acc. to CGA V5-1989 or
3)
N-F
S90-116
WARNING
Only use medical grade oxygen.
WARNING
Only use compressed gas cylinders and pres-
sure reducers, which comply with all applica-
ble regulations and have been approved.
Must have a blow-off valve on the output side to limit
the outlet pressure to approximately 1000 kPa in
the event of a fault.
Average 0.5 L/min
15 % or ±1 L/min, whichever is greater.
Instructions for use Oxylog 3000 plus SW 1.n