Inspiratory and expiratory resistance of
the device with pediatric breathing circuit
Nominal flow
Dead space including flow sensor, but
excluding accessories such as filter, HMEs
and CO
cuvette
2
Dead space of the
CO
cuvette
2
Supplementary functions
Patient connection
1) At a supply pressure >350 kPa.
The maximum inspiratory flow is reduced to 80 L/min at a supply pressure of <350 kPa, and to 39 L/min at a supply pres-
sure of <280 kPa.
2) When using an accessory with a female connector, add 2 mL to the breathing circuit dead space.
Instructions for use Oxylog 3000 plus SW 1.n
Disposable breathing
circuit for adults,
1.5 m
Disposable breathing
circuit for adults,
3.0 m
Reusable breathing
circuit for adults,
1.5 m
Reusable breathing
circuit for adults,
3.0 m
Disposable breathing
circuit for pediatric
patients
Demand valve
Relief valve
<5 mbar at 30 L/min
<2 mbar at 5 L/min
Inspiratory 21 L/min at 2 mbar
Expiratory 10 L/min at 2 mbar
Inspiratory 20 L/min at 2 mbar
Expiratory 10 L/min at 2 mbar
Inspiratory 22 L/min at 2 mbar
Expiratory 16 L/min at 2 mbar
Inspiratory 21 L/min at 2 mbar
Expiratory 15 L/min at 2 mbar
Inspiratory 12 L/min at 2 mbar
Expiratory 7 L/min at 2 mbar
2)
Approx. 35 mL
(reusable breathing circuit for adults).
2)
Approx. 30 mL
(disposable breathing circuit for adults).
2)
Approx. 15 mL
(disposable breathing circuit for pediatric patients).
Approx. 4 mL (CO
cuvette)
2
Approx. 1.5 mL (CO
cuvette for pediatric patients)
2
Opens the breathing system upon failure of the gas
supply, permits spontaneous breathing with ambient
air.
Opens the breathing system in case of a device mal-
function to approximately 80 mbar.
22 mm ISO conical connector
Technical data
143