Dräger Oxylog 3000 plus Instructions For Use Manual page 15

Emergency and transport ventilator software
Hide thumbs Also See for Oxylog 3000 plus:
Table of Contents

Advertisement

WARNING
Do not use the device in conjunction with
magnetic resonance imaging (MRI, NMR, NMI).
The medical device may malfunction, causing
danger to the patient.
WARNING
Risk of explosion and fire
This device is not approved for use in areas
where oxygen concentrations above 25 Vol%
or combustible or explosive gas mixtures are
likely to occur.
WARNING
In toxic surroundings:
– The patient must be ventilated with 100 %
medical grade oxygen so that toxic ele-
ments do not enter into the breathing gas.
– The patient must be immediately trans-
ferred to a breathable atmosphere in order
to prevent inhalation of toxic air when
spontaneous breathing resumes.
WARNING
In infectious environments:
– The patient must be ventilated with 100 %
medical grade oxygen so that bacteria,
viruses, fungi, or spores do not enter the
breathing gas.
– The patient must be immediately trans-
ferred to a breathable atmosphere in order
to prevent inhalation of infectious air when
spontaneous breathing resumes.
Instructions for use Oxylog 3000 plus SW 1.n