External power supply

External power supply with
DC/DC converter
WARNING
The DC/DC converter should be used in dry
locations only.
Risk of electric shock and risk of damage to
device.
The DC/DC converter must be used to connect
Oxylog 3000 plus to on-board power supply
systems, e.g., in emergency vehicles. It can be
used with the following voltages: 12 VDC, 24 VDC
or 28 VDC. The on-board power supply must be
protected by a 10 to 16 A DC fuse.
Oxylog 3000 plus cannot use the DC input power
outside of this range.
Mount the DC/DC converter on a flat wall and make
sure the wall is solid enough to support the bracket.
Use all four mounting holes (screw size M4).
Instructions for use Oxylog 3000 plus SW 1.n
Oxylog 3000 plus
D
C
1 Plug the large connector (A) of the
DC/DC converter into the on-board supply.
2 Plug the small connector (B) into the
DC socket (C) of Oxylog 3000 plus.
3 When Oxylog 3000 plus is correctly connected
to an external power supply, the indicator
(D) lights up.
Assembly
12 VDC
24 VDC
28 VDC
A
B