Reprocessing
Removing the reusable breathing circuit for
adults
1 Disconnect the breathing hose (A) from the
expiratory port.
2 Detach the flow measuring lines (B) from the
connectors.
122
B
A
Disassembling the reusable breathing circuit
for adults
A
C
1 Disconnect the flow sensor (A) from the
breathing valve.
2 Carefully detach the flow measuring lines (B)
from the flow sensor.
Pull the flow measuring lines straight off the
connectors.
CAUTION
Do not twist or use force when disconnecting the
flow measuring lines from the flow sensor noz-
zles.
Otherwise the flow sensor may be damaged.
3 Detach the angled connector (C) from the flow
sensor.
4 Detach the breathing hose (D) from the
breathing valve.
Instructions for use Oxylog 3000 plus SW 1.n
D
B
B