Verifying Back Office Connection - Voc - GE LOGIQ S8 Service Manual

Hide thumbs Also See for LOGIQ S8:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5394227, 12
IRECTION

3-14-3 Verifying Back Office Connection - VOC

To enable disruptive mode or VCO, the user in front of the scanner has to have Remote service access
rights. Otherwise the request for disruptive mode will be automatically denied.
1.)
On Console, invoke "Disruptive Mode" and "VCO" by left click "GE ICON" and select "Connect
Clinical Lifeline".
2.)
Login to FFA Site from Remote PC using Chrome web browser. As of 2018, FFA Site address is
https://ffa.am.health.ge.com.
3.)
In System ID field, type the CRM Number of the machine and select [Get Started].
4.)
At CONNECT mode, Select [Connect] of the Virtual Console on Remote Console..
5.)
After selecting [Connect] button, operator may need to wait long time to launch the remote service
desktop.
6.)
Verify console desktop can be viewed from Remote PC.
3 - 94
PROPRIETARY TO GE
Section 3-14 - Agent Configuration (R4.2.5x and later)
LOGIQ S8/LOGIQ E8 S
M
ERVICE
ANUAL