GE LOGIQ S8 Service Manual page 406

Hide thumbs Also See for LOGIQ S8:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5394227, 12
IRECTION
9-8-11
Mains on applied part (cont'd)
Where applicable, a typical test setup of non-TEE Probes can be as illustrated in
current test" on page
A typical test setup for TEE probes could be as indicated below:
The handle of the TEE probes must not be immersed
WARNING
WARNING
The test passes when the reading measure less than the values in: 9-13 on page 9-19.
9 - 32
9-28.
Figure 9-14 TEE Probe Leakage Isolation (Sink) Current Test
Section 9-8 - Electrical Safety Tests
LOGIQ S8/LOGIQ E8 S
9-8-10 "Probe leakage
M
ERVICE
ANUAL