Grounding continuity

DANGER
LOGIQ V2/LOGIQ V1 – Basic Service Manual
5610739-100 English Rev.10
ELECTRIC SHOCK HAZARD. THE PATIENT MUST NOT BE
CONTACTED TO THE EQUIPMENT DURING THIS TEST.
Measure the resistance from the third pin of the attachment plug
to the exposed metal parts of the case. The ground wire
resistance should be less than 0.2 ohms. Reference the
procedure in the IEC60601-1.
1. GROUND PIN
2. OHMMETER
3. LOGIQ V2/LOGIQ V1
4. ACCESSIBLE METAL PART:
• MONITOR HOUSING
• PEAR PANEL CONNECTOR
• ANY CASTER/WHEEL SUPPORT
Figure 10-3. Ground continuity test
Electrical safety tests
10-25