Setup the LOGIQ V2/LOGIQ V1 for DICOM Communications
(continued)
Table 3-12:
1
Select MyComputer.
2
Assign an AE title to the
DICOM services use this to identify the
case-sensitive. This title may contain the Computer Name from the TCP/IP page,
if desired.
Note: It is NOT recommended to use the factory default. This is not prohibited, but
more than one system with the same AE title can cause confusion.
3
Edit Port Number if needed. 104 is typical. Save your changes and reboot the
system.
LOGIQ V2/LOGIQ V1 – Basic Service Manual
5610739-100 English Rev.10
Utility ->Connectivity ->Device screen
. (AE stands for Application Entity.
LOGIQ V2/LOGIQ V1
LOGIQ V2/LOGIQ V1
Connectivity setup
.) AE title is