Power ON/Boot Up

Warnings
DANGER
DANGER
DANGER
CAUTION
CAUTION
LOGIQ V2/LOGIQ V1 – Basic Service Manual
5610739-100 English Rev.10
ALWAYS CONNECT THE ULTRASOUND SYSTEM TO A
FIXED POWER SOCKET WHICH HAS THE PROTECTIVE
GROUNDING CONNECTOR.
NEVER USE A THREE-TO-TWO PRONG ADAPTER; THIS
DEFEATS THE SAFETY GROUND.
ENSURE THAT THE POWER CORD AND PLUG ARE
INTACT AND THAT THE POWER PLUG IS THE PROPER
HOSPITAL-GRADE TYPE (WHERE REQUIRED).
Ultrasound system requires all covers.
Operate this Ultrasound system only when all board covers and
frame panels are securely in place. The covers are required for
safe operation, good Ultrasound system performance and
cooling purposes.
Use only power supply cords, cables and plugs provided by or
designated by GE.