Loading the System Software with USB memory stick (For R1.1.x)
NOTE:
NOTE:
BIOS Setting
NOTE:
NOTE:
LOGIQ V2/LOGIQ V1 – Basic Service Manual
5610739-100 English Rev.10
This section describes how to upgrade the software of LOGIQ
V2/LOGIQ V1 from R1.0.x to R1.1.x and software upgrade
between R1.1.x.
Load the disk completely to do this software upgrade from
R1.0.x to R1.1.x. When loading the complete disk, ensure that
any patient data on the disk has been backed up.
Patient Archive, Report Archive, User Defined Configuration and
Service of R1.0.x are not compatible with R1.1.x. Do not restore
them between R1.0.x and R1.1.x.
To upgrade the software from R1.0.x to R1.1.x, BIOS setting is
required before software upgrade procedure.
1. Power on the system. Press DEL or F2 immediately when
the BIOS information screen displays. Record the CPU
GPN number.
Figure 7-10. BIOS Revision Information
The GPN number in the figure is for illustrational purpose
only. Refer to your actual output screen on your system for
the correct GPN number.
Loading the software